| Model Number |
MDLR CNLTD TRIDRV DRVR |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported that during the case, the surgeon was drilling, and a pin fell out of the drill and onto the floor.No parts or pieces fell into the patient wound site.The event did not effect the surgical outcome.Product is available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
After review of additional information received the following sections, have been updated accordingly: h6: the broken tri-drive modular handle reported in case-2022-00006554 may have been due to numerous surgical uses and subsequent sterilization cycles, which resulted in the dowel pin disassociating from the assembly.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
|
| |
|