| Model Number |
030403 |
| Medical Device Problem Codes |
Break (1069); Failure to Fire (2610); Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a pneumothorax procedure, the stapler did not worked, the stapling could not be completed.The second handle did not work as well.It was noted that the tissue was too thick for the purple reload.To resolve the issue, a third handle of a different type was used.There was no patient injury.Pli30 - medtronic's evaluation of the device found that visual inspection of internal components revealed sheared teeth on the firing rack.
|
| |
|
Additional Manufacturer Narrative
|
|
Product id: 030403 (030403 gia universal for open surgery, lot number: p1j1399); sigtrsb60amt (sigtrsb60amt 60mm med thk reinforced rel, lot number: unknown).Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the instrument loaded, clamped, yet would not cycle due to sheared firing rack teeth damage.It was reported that the device did not work as expected.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur in any of the following circumstances: 1.Firing over tissue that is beyond the recommended thickness range.2.Firing with an obstacle incorporated in the jaws.3.Attempting to fire a reload that is in interlock.In each circumstance, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a pneumothorax related to bullous emphysema and hemothorax via video thoracoscopic procedure, in resection of apical dystrophy, the clamp broke when the staple application was not complete.This resulted in incomplete stapling without cutting the pulmonary parenchyma.The surgeon had to cut with scissors between the staples (applied but not cut due to the clamp breakage) to be able to apply a new staple without it being bothered by the metal of the staples from the previous line.The second handle of the same type and batch was used but did not make the second staple line.It was noted that the tissue was too thick for the purple reload.To resolve the issue, a third handle of a different type was used.The operation time was extended for abut 15 minutes.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, e1(facility name, street 1, city), g3, h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|