• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number TRX-800
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture, anxiety/textured.
 
Event or Problem Description
Patient reported left side rupture, exchange from textured due to concern with product, and "achy".Patient also reported the non device related events of "hand tingling", "sinus problems, hair loss", and "allergies".The device remains implanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, b6, d6b, d9, h6.
 
Event or Problem Description
The device has been explanted.
 
Event or Problem Description
Patient reported left side rupture, exchange from textured due to concern with product, and "achy".Patient also reported the non device related events of "hand tingling", "sinus problems, hair loss", and "allergies".Patient additionally reported "cyst", and "allergic reaction/hypersensitivity." the device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, broken on the radius side assessed as unidentified (tear) opening.(shell thickness was within specification).Anxiety-product/procedure: unable to observe since it is a medical event and is not related to the device.Pain: unable to observe since it is a medical event and is not related to the device.Sensation increase/decrease-ndr: not applicable as the event is not related to the device.Varied injuries-ndr: not applicable as the event is not related to the device.Additional observations: red particle observed on the gel.Red particle observed on the device.No further actions are required as the device was implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15744313
Report Number9617229-2022-19603
Device Sequence Number17474990
Product Code FTR
UDI-Device Identifier10888628008632
UDI-Public10888628008632
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/13/2021
Device Model NumberTRX-800
Device Catalogue NumberTRX-800
Device Lot Number2868752
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer01/12/2023
02/08/2023
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date01/31/2023
03/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/13/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient SexFemale
Patient Weight84 KG
-
-