• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON ET TUBE, SHER-I-BRONCH, LS, 35 FR; TUBE, TRACHEAL/BRONCHIAL, DIFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON ET TUBE, SHER-I-BRONCH, LS, 35 FR; TUBE, TRACHEAL/BRONCHIAL, DIFF Back to Search Results
Model Number IPN048995
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.Verification of failure mode reported in the current manufacturing process could not be performed due to product was not being manufactured.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported on (b)(6)2022, that the y-tube broke off during use on a patient.No reports of patient having desaturations during the time of event.No patient injury or consequence was reported, and no medical intervention was needed.The current patient's condition was reported as fine.
 
Additional Manufacturer Narrative
Qn#(b)(4).The customer returned one unit 5-16035 et tube, sher-i-bronch, ls, 35 fr for investigation.The et tube, y connector, and both swivel connectors were returned out of its packaging.Two suction catheters were returned in their original packaging.The returned connectors were visually examined with and without magnification.Visual examination of the returned samples revealed that the bronchial swivel connector was broken on the 15mm side.The swivel connector is a component purchased from a supplier.A capa was previously opened to further investigate this issue.Corrective actions were implemented under the capa on 15apr2022 to prevent this issue from recurring.This sample was manufactured prior to the corrective actions.
 
Event or Problem Description
It was reported on (b)(6) 2022, that the y-tube broke off during use on a patient.No reports of patient having desaturations during the time of event.No patient injury or consequence was reported, and no medical intervention was needed.The current patient's condition was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON ET TUBE, SHER-I-BRONCH, LS, 35 FR
Common Device Name
TUBE, TRACHEAL/BRONCHIAL, DIFF
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15744346
Report Number3003898360-2022-00631
Device Sequence Number5850531
Product Code CBI
UDI-Device Identifier14026704617392
UDI-Public14026704617392
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K180253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/07/2026
Device Model NumberIPN048995
Device Catalogue Number5-16035
Device Lot Number73C2100179
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-