• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CONFIRM; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM3500
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2022
Event Type  malfunction  
Event Description
It was reported that a patient presented to the clinic on (b)(6) 2022 for a routine end of service insertable cardiac monitor (icm) explant.During the explant, it was noted that the header of the device almost broke off.The device was explanted without further consequences.The patient was stable throughout.
 
Manufacturer Narrative
The reported event of detachment of device or device component was confirmed.Device was received with header detached from can.Visual inspection found tool mark on device¿s can and broken header which is consistent with explant damage.Longevity assessment was performed, and device had exceeded the projected longevity.As a result of this finding, abbott is performing further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15744709
MDR Text Key303171210
Report Number2017865-2022-44502
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/04/2021
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberP000100685
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/21/2022
Initial Date FDA Received11/07/2022
Supplement Dates Manufacturer Received12/20/2022
01/09/2023
Supplement Dates FDA Received01/08/2023
01/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-