This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Reporter is a j&j sales representative.Investigation summary: the complaint devices are not being returned, they are implanted, therefore unavailable for a physical evaluation.However, a photo was provided.Upon visual inspection of the photo, is showing the loop of suture which should be passing through the holes in the anchor, pulled out, so the anchor broke allowing the suture to separate from it.; therefore, this complaint can be confirmed.A manufacturing record evaluation was performed for the finished device lot number: 7l47939, and no nonconformances were identified.The photo does not provide enough evidence to determine root cause.Without physical evaluation of the device reported, we cannot establish a root cause for the issue experienced by the customer.The possible root cause can be associated with such handling of the device or product interaction during procedure.However, it cannot be conclusively affirmed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|