• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® TIGERTAIL¿ FLEXIBLE TIP URETERAL CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® TIGERTAIL¿ FLEXIBLE TIP URETERAL CATHETERS Back to Search Results
Model Number 139005
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that there was a pinch near the non marked end in few of the polyurethane ureteral catheters that made it hard to pass a wire.Per follow-up via email on 04nov2022, it was reported that they opened two tiger tails that were like this.The third one was okay.
 
Event or Problem Description
It was reported that there was a pinch near the non marked end in few of the polyurethane ureteral catheters that made it hard to pass a wire.Per follow-up via email on 04nov2022, it was reported that they opened two tiger tails that were like this.The third one was okay.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned and further investigation was not conclusive.Although a specific root cause cannot be determined, potential root cause for this event could be, "device was damaged during handling".The device was attempted to be used for treatment purposes.It is unknown if the device had met relevant specifications or resulted in the reported event.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "do not use if package is damaged".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® TIGERTAIL¿ FLEXIBLE TIP URETERAL CATHETERS
Common Device Name
URETERAL CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15745879
Report Number1018233-2022-08519
Device Sequence Number5851081
Product Code KOD
UDI-Device Identifier10801741127097
UDI-Public(01)10801741127097
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K033719
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number139005
Device Catalogue Number139005
Device Lot NumberREGU2450
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2022
Supplement Date Received by Manufacturer03/15/2023
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date03/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-