| Model Number |
71675513 |
| Medical Device Problem Codes |
Fracture (1260); Migration (4003)
|
| Health Effect - Clinical Codes |
Pain (1994); Implant Pain (4561)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, after surgery performed on may, the patient experienced pain, it was found that the nail had come off.It was decided to wait, 1 week later the patient had more pain, the x-ray then showed that the compression screw has come off.This adverse event was treated by a reoperation and patient received a new nail on
(b)(6) 2022.Current health status of patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Section b5 and h6 were updated.Section h10: the devices were not returned for evaluation.Based on the information attached in the complaint and tracking number provided the devices have been in transit since 29/nov/2022; shipper stated the carrier is investigating.However at this time we have no evidence that it was ever received by the complaint investigation team for evaluation, therefore this case will be handled as no product return.If the devices or new information is received, this complaint will be re-opened.The clinical/medical investigation concluded that, the provided x-rays confirm the reported failure mode of the intertan nail and compression screw as well as persistent femoral fractures.However, based on the limited information provided, clinical root cause of the reported events cannot be concluded.The patient impact beyond the pain and revision surgery could not be determined.No further clinical/medical assessment can be rendered at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for trigen intertan 10s 10mm x 38cm 125degree left over the past 12 months and for its batch number based on historical data of the device did not reveal similar events for the listed device.A review of complaint history revealed similar events for intertan lag/compression screw kit 95mm / 90mm over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for intramedullary nail system revealed that proper size, length, side and type selection as well as use of intramedullary nails is essential to safe and effective fracture treatment.Care prior to bony union stablish to immobilize and/or externally support skeletal structures that have been implanted with surgical metallic implants until skeletal union is observed.Early weight bearing substantially increases implant loading and increases the risk of loosening, bending or breaking the device.Early weight bearing should only be considered where there are stable fractures with good bone-to-bone contact.Patients who are obese and/or noncompliant, as well as patients who could be pre-disposed to delayed or non-union, should have auxiliary support.Patients and nursing care providers should be advertised of these risks.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.According with inspection procedure for trigen intertan 10s 10mm x 38cm 125degree left, the inspection includes the verification of part configuration per print.Also, material specification shall control the quality and manufacture of standard grade titanium 6al-4v alloy.According with inspection procedure for intertan lag/compression screw kit 95mm / 90mm, the inspection includes the verification of part configuration per print.Also, material specification shall control the quality and manufacture of standard grade titanium 6al-4v alloy.At this time, we have no reason to suspect that the products failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, size selected or postoperative care.Based on this investigation, the need for corrective action is not indicated.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, after surgery performed on (b)(6) 2022, the patient experienced pain, it was found that the nail had come off.It was decided to wait, 1 week later the patient had more pain, the x-ray then showed that the compression screw has come off.This adverse event was treated by a reoperation and patient received a new nail on (b)(6) 2022.Current health status of patient is unknown.Upon clinical investigation, it was determined that the intertan nail was fractured.
|
| |
|