| Catalog Number |
190-000 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The remainder of the investigation remains in progress.A supplemental report will be provided after completion.Device discarded, single use device.
|
| |
|
Event or Problem Description
|
|
The customer reported a false positive result with the id now covid-19 assay performed (unknown date) on a nasal swab sample collected on (b)(6) 2022.Confirmation testing (platform: (b)(6) lab) on a nasopharyngeal sample collected on (b)(6) 2022 in viral transport media generated a negative result.Per the customer, no additional information will be provided.
|
| |
|
Event or Problem Description
|
|
The customer reported a false positive result with the idnow covid-19 assay performed (unknown date) on a nasal swab sample collected on (b)(6) 2022.Confirmation testing (platform: core lab) on a nasopharyngeal sample collected on (b)(6) 2022 in viral transport media generated a negative result.Per the customer, no additional information will be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Testing was performed in duplicate at abbott diagnostics scarborough, inc.On retained kit lot 1086836 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for kit part number 190-000 / lot 1086836 and test base part number 190-430 / lot 1086836.Quality control release testing met specifications and there were no false negative results observed.The review of the complaints reported as false positive results (confirmed, unconfirmed and conflicting results) related to kit lot 1086836 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue as the logfiles were not provided.A1 (patient identifier) h3 other text : device discarded, single use device.
|
| |
|