• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO HANDPIECE THUMBERSCREW; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO HANDPIECE THUMBERSCREW; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR100026
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during the set-up for navio assisted surgery, a navio handpiece thumberscrew was noticed fractured.Then, when using a back-up screw, it was not possible to insert it to the navio handpiece tracker array.Surgery was performed with a competitor's navigation system.No delays or patient injuries reported.
 
Additional Manufacturer Narrative
H3, h6: the navio handpiece thumberscrew, part number pfsr100026, intended for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release a serial number, lot number, part revision or software version is required to link the device to a dhr or nc investigation.Care and caution should be exercised during the surgical site setup, surgery and tear down to protect the device from an unforeseen force, such as falling onto a hard surface or damage.Refer to the navio surgical system user¿s manual for proper handling.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with repeated over torquing resulting in thumbscrew shaft fracture.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE THUMBERSCREW
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15748370
Report Number3010266064-2022-00647
Device Sequence Number17721441
Product Code OLO
UDI-Device Identifier00885556648421
UDI-Public00885556648421
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160537
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPFSR100026
Device Catalogue NumberPFSR100026
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Supplement Date Received by Manufacturer12/04/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date12/07/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NAVIO SURGICAL SYSTEM JAPAN, SN:(B)(6).
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-