Correction: d9/h3, h6 method, results, conclusion codes.No problem could be observed on the nail adapter t2 ankle, the device is still fully functional.However, it could be confirmed that the targeting arm is deformed and is not functional anymore.Therefore the reported event could be confirmed.The device inspection revealed the following: the identification of the nail adapter was confirmed based on the catalog # and the lot # marked.The device shows normal signs of usage, as scratches and material wear could be observed in the area where the nail adapter is in contact with the targeting arm during assembly, especially the nail adapter "tongue", which is worn out.The reported use of the mallet in order to assemble the instruments contributed to the damage on the ¿tongue¿.It is not possible to determine if the device was also damaged during previous surgeries.Targeting arm: local material sizing due to deformation and wear was observed on the top of one of the lateral locking grooves.The material sizing led to a local decrease of the groove width.The deformation and wear observed were most probably caused by mishandling of the device, such as hammering or excessive torsional load applied with the nail adapter during previous use.It is not possible to determine if the deformation observed was caused during previous usage or if it was fully caused by the mallet used during the surgery in order to assemble the instruments, as reported in the complaint.However, it can be stated that this may have contributed to damage the targeting arm.The returned targeting arm and nail adapter were used to simulate the assembling and disassembling, as performed during a surgery.For one lateral locking groove and for the posterior locking groove, the assembling and disassembling were performed as intended.However, the nail adapter could not be inserted into the second lateral locking groove, where the material sizing was observed.While the adapter is still fully functional, the functionality of the targeting arm could not be given anymore.Thus, the reported event could be confirmed.The dimensional inspection has shown that the nail adapter is according to specification.The bottom of the deformed lateral locking groove is according to specification, while the top of the groove is out of specification.This gives indication that the local decrease of width occurred most probably during the use of the device and is not related to manufacture.Based on investigation, the root cause was attributed to an user related issue.The material deformation was most probably caused by improper use of the targeting arm, such as hammering.As a reminder, the instructions for use clearly state that: "do not hit instruments unless they are specifically intended for impaction.Never hit targeting devices or elastosil handles other than those with an integrated strike plate! always treat the instrument carefully to avoid surface damage or alterations to the geometry." a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
|