TORNIER S.A.S. TORNIER FLEX SHOULDER SYS REV TRAY 12TH 3.5MM OFFSET; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number DWF522
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event or Problem Description
It was reported that there was luxation of the implant after rehabilitation and the patient underwent a revision surgery.
 
Event or Problem Description
It was reported that there was luxation of the implant after rehabilitation and the patient underwent a revision surgery.
 
Additional Manufacturer Narrative
Please note the correction to the d4 lot#: the reported event could be confirmed since an x-ray was provided and matches the alleged failure mode.Indeed, since a post-operative x-ray was provided, the opinion of the medical expert was requested and stated: "without further x-rays and clinical and technical information, it is difficult to make a detailed assessment.The provided x-ray shows indeed a disassembly / dissociation between the aequalis ascend flex stem and the reversed tray [tray completely out of the stem].Therefore, the normal alignment between the humerus and the glenoid is lost (joint dislocation).The reversed tray is 12 mm thick, which is exceptional.I cannot determine the root cause on the information available at this time".The devices inspection revealed the following: -visually, some damage portions and scratches were observed at the flat surface level of the stem, and deep scratches were observed at the taper level of the tray (damages probably due to frictions of the tray on the stem, performed when the tray disassembled and migrated out of the stem in-vivo).-dimensionally, all devices met the specifications at the functional tray/stem portion.-functionally, a tray/stem assembly was carried out and showed that the two devices could assembled with no issue to report.But due to the nature of the returned devices, it is not possible to guarantee the impaction at 100% reliable.Unfortunately, no immediate post-operative surgery x-rays were provided for investigation to check if the tray was indeed correctly seated into the stem immediately after surgery.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.As a reminder, the labeling documents state that: "the surgeon should inspect the implant tapers and mating surfaces for debris or blemishes before assembly.The tapers should be clean and dry for assembly".Moreover, "the following situations threaten the success of the shoulder replacement implant: - sport activity or high activity level (.)" and "the fact that active motion in the arm is restricted in daily activity as only elbow, wrist and finger motion is allowed during the rehabilitation (.)".No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER FLEX SHOULDER SYS REV TRAY 12TH 3.5MM OFFSET
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15749186
Report Number3000931034-2022-00364
Device Sequence Number15008981
Product Code KWS
UDI-Device Identifier03700434019322
UDI-Public03700434019322
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K122698
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/09/2023
Device Model NumberDWF522
Device Catalogue NumberDWF522
Device Lot Number5960AT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-