• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH STANDARD INSIGNIA COLLARED HIP STEM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH STANDARD INSIGNIA COLLARED HIP STEM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 7000-5503
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 08/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Right greater trochanter fracture intraoperative.Initially being managed conservatively.Scheduled orif possible revision tha (b)(6) 2022.Update: "bone observed to be fractured by one of the investigators (b)(6) 2022 via x-ray imaging.No instruments or implants were in contact with the bone when it was observed to be fractured.X-ray imaging occurred to conclude fracture at emergency department (b)(6) 2022.The revision surgery has been cancelled and subject is being treated conservatively and being rechecked in 3 weeks for further evaluation.".
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.
 
Additional Manufacturer Narrative
Reported event: an event regarding periprosthetic fracture involving a other hip stem was reported.The event was confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: a review of the provided medical documents concluded: "it is confirmed that the patient experienced a spontaneous greater trochanteric about a cementless tha.There was no fall or trauma.The root cause of this fracture cannot be established with this limited documentation." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the provided medical documents concluded: "it is confirmed that the patient experienced a spontaneous greater trochanteric about a cementless tha.There was no fall or trauma.The root cause of this fracture cannot be established with this limited documentation." the exact cause of the event could not be determined.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
Right greater trochanter fracture intraoperative.Initially being managed conservatively.Scheduled orif possible revision tha (b)(6) 2022.Update: "bone observed to be fractured by one of the investigators (b)(6) 2022 via x-ray imaging.No instruments or implants were in contact with the bone when it was observed to be fractured.X-ray imaging occurred to conclude fracture at emergency department (b)(6) 2022.The revision surgery has been cancelled and subject is being treated conservatively and being rechecked in 3weeks for further evaluation." update:(b)(6) 2022 "the revision surgery has been cancelled and subject is being treated conservatively and being rechecked in 3weeks for further evaluation." update: op note date (b)(6) 2022 "subject is doing good, no revision surgery warranted at this time.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDARD INSIGNIA COLLARED HIP STEM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15749272
Report Number0002249697-2022-01593
Device Sequence Number6364052
Product Code JDI
UDI-Device Identifier07613327511413
UDI-Public07613327511413
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K211703
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number7000-5503
Device Catalogue Number7000-5503
Device Lot Number91723203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer01/04/2023
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date01/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
Patient SexFemale
Patient Weight89 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-