Model Number TRX-800
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 09/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: analysis of the returned device identified: weight to spec, green particles in the shell, crease fold and wear abrasion.Cloudy and air voids in the gel observed after autoclave cycle.Base on the device analysis the final assessment is: no issues found related with the manufacturing process.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown and recall.
 
Event or Problem Description
Healthcare professional reported right side "capsular contracture" baker grade unknown and "recall".Device has been explanted and replaced.
 
Event or Problem Description
Healthcare professional additionally reported right "partial double capsule".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15749296
Report Number9617229-2022-09186
Device Sequence Number15008984
Product Code FTR
UDI-Device Identifier10888628008632
UDI-Public10888628008632
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model NumberTRX-800
Device Catalogue NumberTRX-800
Device Lot Number2895742
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer01/04/2023
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date01/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
Patient Weight57 KG
-
-