| Model Number |
ZA9003 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the lens was partially inserted and removed during the same procedure, as the trailing haptic broke off in the patient¿s right eye.There was a 5-minute delay in the procedure.There were no surgical and/or medical interventions and no patient injury reported.The patient has fully recovered.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Give date: n/a (not applicable).There is no indication that the intraocular lens was implanted.If explanted, give date: n/a (not applicable).There is no indication that the intraocular lens was implanted.Therefore, lens was not explanted.Device evaluated by mfr: other (81): the lens was not returned for evaluation as it remains implanted in the eye.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: nov 7, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received inside of the original lens case.The original folding carton, patient stickers, and a folding cartridge were received as well.Additional information on a note card was received as well.Visual inspection under magnification revealed that the complaint lens was receive cut in half and with a haptic detached.The lens was cleaned and no further issues were identified with the lens.Based on the return condition of the lens, no further product evaluation could be performed.Conclusion: the complaint issue "haptic detached" was observed during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: additional information from the account indicated that an emerald cartridge was used, but they did not have the lot number.Additionally, it was stated that the procedure was completed with the backup za9003, 25.0 diopter.The patient is white, 380lbs.The following fields were unpdated accordingly: b5: the procedure was completed with the backup za9003, 25.0 diopter.A4: patient weight: 380lbs.A5: race: white.D10: concomitant medical products and therapy dates: emerald cartridge, lot unknown.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|