Catalog Number 2C8858
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that three (3) paclitaxel sets leaked.The events occurred in the pharmacy during preparation.The leaks were observed from the tubing under the drip chamber.No patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Device manufacturer address 1: (b)(4) parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
H10: one actual device was received for evaluation.The other two (2) samples were not received and therefore, could not be evaluated.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional testing was performed including clear passage and pressure testing which revealed a leak between the assembly of the tubing into the drip chamber.Additional priming was performed at the set with a bag of sterile water which revealed a leak in the same area.The assembly of the tubing into the drip chamber was not according to specifications; low assembly was detected.The reported condition was verified.The cause of the condition was due to assembly issue during the manufacturing process.A batch review was conducted on both of the suspect lot numbers and there were no deviations found related to this reported condition during the manufacture of these lots.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK SOLUTION ADMINISTRATION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key15749671
Report Number1416980-2022-06056
Device Sequence Number6364563
Product Code FPA
UDI-Device Identifier00085412071688
UDI-Public(01)00085412071688
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K160007
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number2C8858
Device Lot NumberR22E05147
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer11/24/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-