• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72404267
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an inflatable penile prosthesis (ipp) revision surgery in which the pump and cylinders were removed and replaced because the device was pierced.No further information was provided.
 
Event or Problem Description
It was reported that during an inflatable penile prosthesis (ipp) implant surgery the cylinders were pierced, therefore, they were removed and replaced.No patient complications were reported.
 
Event or Problem Description
It was reported that during an inflatable penile prosthesis (ipp) implant surgery the cylinders were pierced, therefore, they were removed and replaced.No patient complications were reported.
 
Additional Manufacturer Narrative
Upon receipt at our quality assurance laboratory, these inflatable penile prosthesis (ipp) cylinders and pump underwent a thorough analysis.The components were visually and microscopically examined, and leak/functionally tested.No anomalies were found with the cylinders nor the pump.Based on the information available and analysis results, the product analysis was not able to identify the cause that could have contributed to the reported clinical observation of material puncture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Common Device Name
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key15749771
Report Number2124215-2022-44578
Device Sequence Number6316947
Product Code FHW
UDI-Device Identifier00878953003757
UDI-Public00878953003757
Combination Product (Y/N)Y
Initial Reporter CountryIT
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/08/2023
Device Model Number72404267
Device Catalogue Number72404267
Device Lot Number1100055915
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/21/2022
Supplement Date Received by Manufacturer10/27/2022
12/19/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date11/18/2022
01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-