| Model Number |
OJR410 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Fisher & paykel healthcare (f&p) are currently in the process of retrieving further information regarding this complaint.We will provide a follow up report upon the completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in sweden reported via a fisher and paykel (f&p) healthcare field representative that the tubing of an ojr410 optiflow junior 2 nasal cannulas were detached from the cannula during patient use.There was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The optiflow junior cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint device was not received at fisher & paykel healthcare (f&p) for investigation.Our investigation was thus based on the information provided by the customer, previous investigations of similar complaints, and our knowledge on the product.Results: the customer reported that the tubing of an ojr410 optiflow junior 2 nasal cannula was detached from the cannula.Conclusion: without the complaint device, we are unable to determine the cause of the reported fault.All optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.Samples are also taken and pull tested to check the tube tensile strength and the glue joint strength at the cannula/tube joint, as well as the swivel grip joint.The subject cannula would have met the required specification at time of production.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 cannula.They also state the following: appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death).Do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).Ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in sweden reported via a fisher and paykel (f&p) healthcare field representative that the tubing of an ojr410 optiflow junior 2 nasal cannulas were detached from the cannula during patient use.There was no patient consequence.
|
| |
|