• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 5MM X 20CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 5MM X 20CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035475200
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Intracranial Hemorrhage (1891); Ruptured Aneurysm (4436)
Date of Event 05/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This is 1 of 2 reports.The subject device is unavailable to manufacturer.
 
Event or Problem Description
It was reported that during coil embolization using the coil (subject device), intraoperative rupture occurred and the patient experienced an intracranial hemorrhage.The patient's outcome was unknown.No other information is available.
 
Additional Manufacturer Narrative
Although the lot number was not provided, the automated manufacturing execution system (mes) has controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Based upon medical review, the harm observed in this complaint is anticipated in nature as per the device risk assessment.As a product related root cause does not apply and the issue is due to a known physiological effect of the procedure and/or patient condition noted within the dfu, product labeling and/or risk documentation files, an assignable cause of anticipated procedural complication will be assigned to this complaint.This is 1 of 2 reports.
 
Event or Problem Description
It was reported that during coil embolization using the coil (subject device), intraoperative rupture occurred and the patient experienced an intracranial hemorrhage.The patient's outcome was unknown.No other information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 SOFT 5MM X 20CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15749955
Report Number3008881809-2022-00554
Device Sequence Number6321308
Product Code HCG
UDI-Device Identifier04546540676511
UDI-Public04546540676511
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberM0035475200
Device Catalogue NumberM0035475200
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/26/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date11/07/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-