• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CONQUEST 40; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CONQUEST 40; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQF75124
Medical Device Problem Codes Difficult to Remove (1528); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiration date: 12/2024).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured at 20 atm.It was further reported that the balloon was allegedly difficult to remove.The procedure was completed by using another device.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured at 20 atm.It was further reported that the balloon was allegedly difficult to remove.The procedure was completed by using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one conquest 40 pta dilatation catheter inserted in a cordis introducer sheath was received for evaluation.During visual evaluation, no kinks were noted to the catheter shaft and no anomalies to the luers, bifurcate or glue fillets.During functional evaluation, the balloon was fully deflated and attempted to be removed from the sheath but was unsuccessful.Bunching was noted proximal to the sheath hub.No further functional testing was performed.Therefore, the investigation is confirmed for the reported sheath removal difficulty as the catheter was received inserted in an introducer sheath and an attempt to the remove it from the sheath was unsuccessful.The investigation is inconclusive for the reported balloon rupture as the balloon was stuck in the sheath and unable to be taken out to be evaluated.A definitive root cause for the reported balloon rupture and sheath removal difficulty could not be determined based upon the provided information.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15750174
Report Number2020394-2022-00872
Device Sequence Number12353369
Product Code DQY
UDI-Device Identifier00801741060663
UDI-Public(01)00801741060663
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQF75124
Device Catalogue NumberCQF75124
Device Lot NumberREGN4198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexMale
Patient Weight78 KG
Patient EthnicityHispanic
-
-