U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NAVIGATOR HD; ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NAVIGATOR HD; ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number M0062502260
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that a navigator hd access sheath device was used during a ureteroscopy procedure performed on (b)(6) 2022.During unpacking, the device was inspected, and it was noticed that the black tip of the dilator was broken.The procedure was completed with another navigator hd access sheath device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a04 captures the reportable event of dilator tip was broken.Investigation result: the returned navigator hd device was analyzed, and a visual evaluation noted that the dilator and the sheath were returned for analysis; however, the detached section was not returned.It was noted that the dilator security tab of the hub was broken.The tip of the sheath and the dilator were found in good condition.Microscopic examination was performed and confirmed under magnification that the dilator security tab of the hub was broken.No other problems with the device were noted.This problem is likely due to factors encountered during the procedure, such as handling and manipulation of the device, and/or excessive force applied to the device during use could have caused the event.Therefore, the most probable root cause is manufacturing deficiency.An investigation to address this problem is in progress.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a navigator hd access sheath device was used during a ureteroscopy procedure performed on (b)(6) 2022.During unpacking, the device was inspected, and it was noticed that the black tip of the dilator was broken.The procedure was completed with another navigator hd access sheath device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIGATOR HD
Common Device Name
ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15751332
Report Number3005099803-2022-06462
Device Sequence Number20384250
Product Code FED
UDI-Device Identifier08714729837558
UDI-Public08714729837558
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K140323
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2023
Device Model NumberM0062502260
Device Catalogue Number250-226
Device Lot Number0028233530
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
Patient Weight80 KG
-
-