|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the cautery tip of the instrument was damaged.The button had disengaged and was damaged.It was reported that the sheath was damaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the user excessively manipulates the button and\or exerts angular pressure on the button which results in damage to the button assembly.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: verify mechanical and electrical compatibility of devices from different manufacturers prior to using them together in a procedure.Always ensure that the outer sleeve is pushed fully forward over the tip of the device prior to insertion or removal of the device through the trocar sleeve.Failure to do so may cause damage to the device.The operator should regularly inspect these devices for possible damage, in particular, the electrode cables.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|