This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: b4, b5, d4, g3, g6, h1, h2, h3, h4, h6, h10. visual examination of the returned product identified signs of repeated use (nicked/gouged) and the device is fractured into two pieces, with all pieces returning.Review of the device history records identified no deviations or anomalies during manufacturing.Insufficient information provided to perform a compatibility check.Medical records were not provided.Complaint is confirmed.The device has a potential field age of over 12 years and exhibit signs of repeated use.The root cause of the reported event is attributed to wear and tear from repeated use over time.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|