This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device running in locked position - power broken, identified during service and evaluation was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.Udi: (b)(4).
|
It was reported that the motor device was not working properly.During in-house engineering evaluation, it was observed that the device ran in locked position- power broken and had foreign substance/debris/cleaning/sterilization.It was further determined that the device failed pretest for motor thermistor assessment and safety assessment.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|