Model Number 100/199
Medical Device Problem Code Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the pre-use check, the cuff was unable to be deflated.No patient injury reported.
 
Additional Manufacturer Narrative
Operator of device serial number/lot number is unknown, no product information has been provided to date.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Other text: the lot number was provided.(updated d4 and h4).H6: event problem and evaluation codes: update not required.
 
Additional Manufacturer Narrative
H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample and two photos were received for evaluation.Functional testing found leakage was detected between the check valve and pilot balloon.The root cause of the reported issue was found to be not enough solvent at joint.Customer complaint notification was performed on (b)(6) 2022 to production personnel as awareness to failure mode reported.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ENDOTRACHEAL TUBES
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
3-10-15, hakata station east,
minneapolis, MN 55442
MDR Report Key15752994
Report Number3012307300-2022-26881
Device Sequence Number2958899
Product Code BTR
UDI-Device Identifier15019315022357
UDI-Public15019315022357
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K790312
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2025
Device Model Number100/199
Device Catalogue Number100/199/085
Device Lot Number3990337
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer11/08/2022
05/30/2023
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date11/14/2022
06/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-