• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 12FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 12FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884715122
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that a 2-month-old patient was transferred to the surgery room on the day to perform a tracheostomy and gastrostomy, during the placement of the #12 fr gastrostomy tube, the balloon was broken.Additional information was received and stated that there were no fragments left inside the patient.Per customer, the patient is in good health.
 
Additional Manufacturer Narrative
A device history record review could not be performed because a lot number was not received with the complaint.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.A sample was not received for the investigation.Per customer, the sample was discarded.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the issue and root cause analysis.At this time, a corrective and preventive action is not deemed necessary.As an awareness notification, the production personnel received an awareness notification.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSTRO FEED TBE W/Y PRT 12FR
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15753009
Report Number9612030-2022-03455
Device Sequence Number2859088
Product Code KNT
UDI-Device Identifier10884521007154
UDI-Public10884521007154
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884715122
Device Catalogue Number8884715122
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date01/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient Age2 MO
Patient SexUnknown
-
-