• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH DOBBHOFF 12FR 43IN W STYLET; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH DOBBHOFF 12FR 43IN W STYLET; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884711253
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Customer reports: the nasoenteric tube was installed at the end of the afternoon shift, but after an x-ray was taken, the tube showed total resistance.The probe was removed for a new procedure and it presented a "bend".This fact has happened other times, a new probe, was passed on the patient and there was no resistance.
 
Additional Manufacturer Narrative
Device evaluation comment - the customer stated that the device will not be returned for evaluation because it was contaminated and discarded.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
A review of the device history record (dhr) revealed no abnormal process conditions were present during the manufacturing of the product that could have led to the condition described by the customer.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.There were no samples received for evaluation therefore the reported condition could not be confirmed, nor could a root cause be determined.A corrective/preventative action plan is not required since the reported condition has not been identified as manufacturing related.The current process is running according to product specifications meeting quality acceptance criteria.We will continue monitoring the process for any adverse trends that require immediate attention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOBBHOFF 12FR 43IN W STYLET
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
MX  02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15753927
Report Number9612030-2022-03456
Device Sequence Number14170269
Product Code KNT
UDI-Device Identifier10884521518803
UDI-Public10884521518803
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884711253
Device Catalogue Number8884711253
Device Lot Number2200306464
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date01/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-