• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC DISPOSABLE CIRCUITS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC DISPOSABLE CIRCUITS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Model Number 100/905/340-CH
Medical Device Problem Code Pressure Problem (3012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the use of the product the customer maxed out the peep (positive end expiratory pressure) setting, but the actual value reached 12-14 cmh2o.No patient injury reported.
 
Additional Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Manufacturing device history record review was not performed because manufacturing records are retained by the supplier at their manufacturing site and are not readily available for review.A product sample was received for evaluation.Visual inspection and functional testing was performed.Visual inspection found no abnormality in the appearance of the product.Functional testing found the positive end-expiratory pressure (peep) value was slightly lower than that of the normal product when the complaint product was connected to the parapack ventilator and operated.The root cause of the reported issue could not be determined.The product sample was returned to the supplier.This issue will be monitored in an internal scar for any increase in occurrence.If the occurrence of this issue increases significantly, additional corrective actions will be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC DISPOSABLE CIRCUITS
Common Device Name
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK  LU3 4BU
Manufacturer (Section G)
NULL
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK   LU3 4BU
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka,
minneapolis, MN 55442
MDR Report Key15754434
Report Number3012307300-2022-26895
Device Sequence Number15032021
Product Code BTL
UDI-Device Identifier15019517160215
UDI-Public15019517160215
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K123957
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/15/2026
Device Model Number100/905/340-CH
Device Catalogue Number100/905/340-CH
Device Lot Number210315
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer12/05/2022
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-