| Model Number |
480257 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number 400574395.The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: multiple alarms of air in line, multiple attempts to fix the issues, 20 minutes delay in treatment.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).One used sample without packaging was returned for evaluation.The sample was visually evaluated with no non-conformance were noted.The sample was attached to observe if it would fit into the pump the tubing did not need to be stretched and there was no evidence of interference between the tubing couplers and the pump housing.In an attempt to replicate the reported defect of the air-in-line alarm and occlusion alarm on the pump, the sample was attached to the pump and then tested by running therapy while staggering the rate of flow throughout the therapy.No alarms occurred during the testing of the returned sample.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|