ROCHE DIAGNOSTICS COBAS 8000 CORE UNIT; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number COBAS 8000 CORE
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service engineer performed flow path cleaning replaced tubing, replaced 2 reagent syringes, and adjusted the sipper nozzle position.An instrument check was performed successfully.A reproducibility test was performed using pooled serum and the results were acceptable.The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable roche diagnostics elecsys anti-tg results for 1 patient sample on a cobas 8000 core unit.The initial anti-tg result was 47 iu/ml.An aliquot from the primary tube was repeated and the result was 93 iu/ml with flag.The primary tube was repeated and the result was 52 iu/ml with flag.Another aliquot from the primary tube was repeated and the result was 53 iu/ml with flag.No questionable results were reported outside of the laboratory.The reagent lot is 59537501 and the expiration date is 28-feb-2023.
 
Additional Manufacturer Narrative
Based on the calibration and qc data, a general reagent issue could be excluded.The investigation determined that the root cause of the event is consistent with pre-analytical handling issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 8000 CORE UNIT
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15754621
Report Number1823260-2022-03547
Device Sequence Number2918557
Product Code JJE
UDI-Device Identifier04015630928330
UDI-Public04015630928330
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCOBAS 8000 CORE
Device Catalogue Number05641446001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date03/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-