| Model Number |
COBAS 8000 CORE |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer performed flow path cleaning replaced tubing, replaced 2 reagent syringes, and adjusted the sipper nozzle position.An instrument check was performed successfully.A reproducibility test was performed using pooled serum and the results were acceptable.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable roche diagnostics elecsys anti-tg results for 1 patient sample on a cobas 8000 core unit.The initial anti-tg primary tube result was 39.6 iu/ml.An aliquot from the primary tube was repeated and the result was 10 iu/ml with flag.The primary tube was repeated and the result was < 10 iu/ml with flag.Another aliquot from the primary tube was repeated and the result was < 10 iu/ml with flag.No questionable results were reported outside of the laboratory.The reagent lot is 59537501 and the expiration date is 28-feb-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available data, a general reagent issue could be excluded.The investigation found that the root cause of the event was consistent with pre-analytical handling issues.
|
| |
|