BAXTER HEALTHCARE CORPORATION RFA, LUER LOCK - LUER LOCK; SYRINGE, PISTON Back to Search Results
Catalog Number H93813901
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a particle was observed inside the rapidfill connectors (red).This was observed prior to use.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
H4: lot manufacture date: july 01, 2020 - july 02, 2020.H10: the device was received for evaluation connected to an unknown 5 ml syringe.Visual inspection was performed and a dark particle matter was observed lodged inside the interior wall of the syringe tip thread area.The sample connector that was attached to an unknown 5 ml syringe was removed and a loose particle matter fell out of the syringe tip area.Visual inspection with magnification verified one dark flake-like polymeric appearing particle approximately 1.00 square millimeters.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFA, LUER LOCK - LUER LOCK
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key15755207
Report Number1416980-2022-06080
Device Sequence Number2859168
Product Code FMF
UDI-Device Identifier00085412478364
UDI-Public(01)00085412478364
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K872743
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2023
Device Catalogue NumberH93813901
Device Lot Number60247257
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-