Model Number 1288010000 |
Device Problems
Smoking (1585); Temperature Problem (3022)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/25/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that there was a thermal event.
|
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
|
Manufacturer Narrative
|
Alleged failure: 1288 smoking in surgery room.It was plugged into a stryker cart.No camera head was plugged in.Did not look like damage to the cart plug.Update - no patient/user injuries.Device was warm to the touch.Heavy gray smoke.] the failure(s) identified in the investigation is consistent with the complaint record.Root cause: the root cause is an electrical component failure on the power supply board.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that there was a thermal event.
|
|