| Model Number |
EVOLUTION 3E T |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
Unspecified Respiratory Problem (4464)
|
| Date of Event |
10/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Event description: the initial reporter located outside the u.S.Reported that while in use on a patient, the evolution ventilator generated low oxygen supply alarm.The patient was removed from the ventilator and placed on an alternate ventilator with no injury reported.Manufacturer narrative: patient information was not provided.The third party service personnel (sp) inspected the ventilator and confirmed the reported issue.Sp replaced the device's oxygen pressure sensor board to address the issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all event medical quality s pecifications.The manufacturer's investigation is ongoing.Event medical will submit a supplemental if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient information was not provided.The third party service personnel (sp) inspected the ventilator and confirmed the reported issue.Sp replaced the device's system daufghter board to address the issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all event medical quality specifications.The manufacturer's investigation is ongoing.Event medical will submit a supplemental if additional information becomes available.
|
| |
|
Search Alerts/Recalls
|