AMO MANUFACTURING NETHERLANDS TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number ZLU150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient information unknown/asked but not provided.Date of event: unknown, not provided, but the best estimate date is between aug 1, 2022 and oct 10, 2022.Device evaluation: the product testing could not be performed as the product was not returned for evaluation.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed and no non-conformance report (nc) was found as part of this manufacturing records review.The product was manufactured and released according to specification.A search in complaint system revealed that no other complaints were received for this production order number.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be verified.Attempts have been made to obtain additional information regarding the event; however, to date, the information has not been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been s submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from the patient's left eye as patient complained of blurry vision.The post-operative (op) vision is 20/20.There were no medical or surgical interventions required.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: dec 8, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received inside a specimen cup.Visual inspection under magnification revealed that the lens was cut in half, which is consistent with a lens that was explanted, and both halves were covered in dry viscoelastic material.The lens was cleaned and further inspected under magnification, and no further issues were observed.No issues that could contribute to the complaint issue were observed.The complaint issue "blurry vision" was not identified during product evaluation.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15756273
Report Number3012236936-2022-02779
Device Sequence Number20057953
Product Code MFK
UDI-Device Identifier05050474689152
UDI-Public(01)05050474689152(17)250506
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/06/2025
Device Model NumberZLU150
Device Catalogue NumberZLU150U090
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-