• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. MDLR CNLTD TRIDRV DRVR; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. MDLR CNLTD TRIDRV DRVR; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number MDLR CNLTD TRIDRV DRVR
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgical use the device broke during reaming.No fragments, parts or pieces fell into the patient.All pieces were accounted for.Patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
After review of additional information received the following sections, have been updated accordingly: h6: the disassembled tri-driver reported was likely the result of the retaining ring responsible for maintaining assembly of the sleeve on the tri-driver shaft becoming deformed as a result of the sleeve being held stationary while reaming or the sleeve being obstructed by soft tissue or bone.The operative technique was updated to warn users against holding the sleeve of the tri- driver stationary while reaming.
 
Additional Manufacturer Narrative
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MDLR CNLTD TRIDRV DRVR
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
ashley harris
3523771140
MDR Report Key15756383
Report Number1038671-2022-01428
Device Sequence Number6675870
Product Code LXH
UDI-Device Identifier10885862085313
UDI-Public10885862085313
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMDLR CNLTD TRIDRV DRVR
Device Catalogue Number315-25-00
Device Lot Number47364013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer02/08/2023
03/24/2026
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date03/06/2023
03/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
Patient SexMale
-
-