Model Number 30914
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device expiration date: unknown.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd alaris¿ extension set experienced leakage.The following information was provided by the initial reporter: leaking.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 05-jan-2023.H6: investigation summary: the customer reported leaking and returned one used set, electrical connection set, and syringe.The sample had an additional short extension set attached, and both the extension set and reported sample were tested for leakage.No leak was found with the sets provided, or with the syringe.It should be noted the extension set was difficult to remove from the reported 30914 set and there was a slight occlusion in the line when first attempted to prime.Neither of these constitute a failure but its worth noting.The reported failure could not be replicated and the complaint could not be verified.The root cause is unknown without an observed failure.A device history record review could not be performed on material 30914 because the lot number is unknown.The root cause of this failure could not be identified without a failure investigation.
 
Event or Problem Description
It was reported that the bd alaris¿ extension set experienced leakage.The following information was provided by the initial reporter: leaking.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS¿ EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15756957
Report Number9616066-2022-01682
Device Sequence Number6239536
Product Code FPA
UDI-Device Identifier50885403233317
UDI-Public(01)50885403233317
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K790108
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number30914
Device Catalogue Number30914
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer01/25/2023
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date02/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-