BARD PERIPHERAL VASCULAR, INC. DENALI FEMORAL SYSTEM; VENA CAVA FILTER Back to Search Results
Model Number DL900F
Device Problems Detachment of Device or Device Component (2907); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2022
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The device has not been returned to the manufacturer for evaluation.However, photos and images were provided for review.The investigation of the reported event is currently underway.
 
Event Description
It was reported that during a vena cava filter placement procedure, the filter was allegedly tilted.It was further reported that the filter legs were allegedly detached, migrated and could be seen in the subsegmental branches of the right pulmonary artery.Reportedly, a snare was used to retrieve the filter.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.Three medical images and two electronic photos were provided and reviewed.The chest and abdominal x-ray of the image shows the filter which is slightly tilted.The filter like object is also found in the right lung field.Based on the image review, the reported positioning problem and detachment issues are confirmed.The provided two photos shows the filter, which is retrieved from the patient.The filter in the photo has at least one leg missing.Hence, based on the photo review, the reported detachment issue is confirmed.Therefore, the investigation is confirmed for the reported positioning problem and the detachment issue.A definitive root cause for the reported positioning problem and detachment issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a vena cava filter placement procedure, the filter was allegedly tilted.It was further reported that the filter legs were allegedly detached, migrated and could be seen in the subsegmental branches of the right pulmonary artery.Reportedly, a snare was used to retrieve the filter.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI FEMORAL SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15757482
MDR Text Key304728452
Report Number2020394-2022-00880
Device Sequence Number1
Product Code DTK
UDI-Device Identifier00801741040801
UDI-Public(01)00801741040801
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDL900F
Device Catalogue NumberDL900F
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexFemale
-
-