• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFAT30
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Type of Reportable Event Serious Injury
Event or Problem Description
A healthcare professional reported overcorrection and lens was explanted in a secondary procedure for a clinical reason of myopia.The lens was replaced with another model lens.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15758201
Report Number1119421-2022-02386
Device Sequence Number6175006
Product Code MFK
UDI-Device Identifier00380652418676
UDI-Public00380652418676
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/23/2026
Device Model NumberTFAT30
Device Catalogue NumberTFAT30.185
Device Lot Number15156664
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/17/2022
Initial Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-