• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIB INSERT IMPL CRC, SZ 2, 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIB INSERT IMPL CRC, SZ 2, 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 02-012-51-2015
Medical Device Problem Codes Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Synovitis (2094); Implant Pain (4561)
Date of Event 10/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this female patient's knee was revised due to pain and synovitis.The surgeon performed releases and implanted a new 15 crc insert with good range of motion and stability.There was no surgical delay or issue with any devices during the revision procedure.The patient left in stable condition.The explant is not available.No additional information is available.
 
Additional Manufacturer Narrative
Based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the pain and subsequent revision cannot be conclusively determined; however, it is most likely related to the patient conditions.Concomitant medical products: (b)(4), 02-010-03-0320 - logic cr femoral cem, right, sz 2.(b)(4), 02-012-45-2020 - lgc tibial fit tray cem sz 2f / 2t.(b)(4), 200-02-35 - three peg patella 35mm.(b)(4), 201-46-10 - holding pin headless 3" (4 pk).Unassigned, 213-49-00 - lpi collar drill.Unassigned, 201-45-00 saw guide holding drill 1/8 x 5.
 
Additional Manufacturer Narrative
If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: b, c.The revision reported was likely the result of tibial insert prosthesis wear.Possible causes of the observed polyethylene wear include third body wear, patient-related factors, inclusion of the polyethylene in the packaging recall or any combination of these possibilities.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIB INSERT IMPL CRC, SZ 2, 15MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
3523771140
MDR Report Key15758396
Report Number1038671-2022-01429
Device Sequence Number6238586
Product Code JWH
UDI-Device Identifier10885862230355
UDI-Public10885862230355
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K123342
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/22/2021
Device Model Number02-012-51-2015
Device Catalogue Number02-012-51-2015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer03/05/2025
09/24/2025
Initial Report FDA Received Date11/09/2022
Supplement Report FDA Received Date03/06/2025
09/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/26/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0021-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-