C.R. BARD, INC. (COVINGTON) -1018233 DAVOL® FEMALE CATH KIT Back to Search Results
Model Number 0035720
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that a patient's nurse was completing in and out of female catheter and the catheter broke while still in the patient.Stated that the nurse was able to hold onto both pieces to avoid the broken piece going up into the patient.
 
Event or Problem Description
It was reported that a patient's nurse was completing in and out of female catheter and the catheter broke while still in the patient.Stated that the nurse was able to hold onto both pieces to avoid the broken piece going up into the patient.Per follow up via email on 07nov2022, it was stated that the catheter broke after placed in patient and both ends of catheter removed safely and patient had to be re-catheterized.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.1 sample were confirmed to exhibit the reported failure.The device had not met specifications.The product was used for patient treatment.The product caused the reported failure.Visual evaluation of the returned photo sample noted one opened (without original packaging), used female catheter.Visual inspection of the photo sample noted the catheter was broken into two pieces.This does not meet specification which states "kit and package must conform to drawings".Although a specific cause cannot be determined, based on the risk document a potential root cause for this event could be "incorrect operation".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "instructions for use: 1.Open package and remove plastic wallet.2.Open plastic wallet and don gloves.3.Pull catheter out of tube to desired length.Lay tube on sterile field.4.Open lubricant and lubricate catheter.5.Open swab packet.Cleanse vaginal area.6.Proceed with catheterization.7.Pull catheter out of top; tighten cover and depress blue spout.8.Fill out label, place on centrifuge tube.Send to lab in normal manner.Important: 1.Use plastic wallet as sterile field.2.Pull catheter out of centrifuge tube to the proper length immediately after donning gloves.Note: if urine does not flow freely into the tube, the cap may need to be loosened slightly.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVOL® FEMALE CATH KIT
Common Device Name
FEMALE CATH KIT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15758548
Report Number1018233-2022-08603
Device Sequence Number18197967
Product Code FFH
UDI-Device Identifier00801741045837
UDI-Public(01)00801741045837
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number0035720
Device Catalogue Number0035720
Device Lot NumberNGGR4121
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer03/13/2023
Initial Report FDA Received Date11/09/2022
Supplement Report FDA Received Date03/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;