| Model Number |
0998-00-3013-58 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the cs100 intra-aortic balloon pump (iabp) does not work.
|
| |
|
Event or Problem Description
|
|
It was reported that the cs100 intra-aortic balloon pump (iabp) does not work.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and was able to reproduce the reported issue.The cs100 displayed a helium failure.The fse discovered the helium bottle was closed.The issue was resolved.The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
| |
|