(b)(4).Event date: only event year known: 2022.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device batch number, and no [non-conformances / manufacturing irregularities] were identified.Photo images were received and are pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.Additional information was requested, and the following was obtained: what is the severity of the burn? (please see degrees of burns below and choose one) first degree burns are minor burns on the first layer of skin.The skin looks dry, redness, and may be swelling; no penetration or blisters.Second degree burn looks wet or moist.The burn site appears red, blistered, and may be swollen and painful third degree burn the burn site looks deep, whitening or blackened and charred - third degree what medical intervention was used to treat the burn? (such as salve or stitches) - antibiotic ointment.Besides the burn, did the patient experience any adverse consequence due to the issue? none documented.Are there any anticipated long-term effects from the burn or injury? not at this time, provider plan is to let burn mature for 1 year and reassess for surgical needs what is the current patient status? stable.Will the device be returning for analysis? if yes, to whom should the shipper kit be sent to? sent to (b)(6) eth/endo surgery.Are the photos of the burn that you could share? if yes please send them to (b)(4).Are the photos of the device that you could share? if yes please send them to (b)(4) - no, sending device.What was the surgical procedure? tonsillectomy.Does the surgeon believe there is there an alleged deficiency to the device that led to patient burn and if so why? yes - outcome.What generator was being used? what power levels was generator set to? what was the estimated activation time of the device? 5 minutes.How was the device cleaned throughout the procedure? manually with sterile cloth.What returned electrode was being used? what pencil was being used? covidien.When the device is returned can we please have the pencil returned as well? what was used to place the electrode into the pencil? manual - hand.What was the exact position of the burn on the lip? see photo.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
(b)(4).Date sent: 12/22/2022.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned samples revealed that the 0118a devices were received in normal wear and with no apparent damaged.The instrument was tested for continuity and no anomalies were noted with the functionality of the device.As part of ethicon endo surgery quality process all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.
|