| Model Number |
IPN028446 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
It was reported that leakage was found between the red and blue fast click connectors while using in the patient.There was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).No sample was returned for analysis.Customer complaint is confirmed, the unit had a leak on ats connector.However, it can't be observe the exactly place where the leak cam from.Product sample is necessary to perform a proper investigation.Nevertheless, a capa 252 was opened in order to address "trend in complaints due to pleur evac leaking".
|
| |
|
Event or Problem Description
|
|
It was reported that leakage was found between the red and blue fast click connector while using in patient.There was no reported injury.
|
| |
|
Search Alerts/Recalls
|