• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. POLYAXIAL SCREW; SIZE Ø5.0X30 MM; POSTERIOR CERVICAL SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. POLYAXIAL SCREW; SIZE Ø5.0X30 MM; POSTERIOR CERVICAL SCREW SYSTEM Back to Search Results
Model Number 7601-05030
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product returned.
 
Event or Problem Description
A company representative reported that the tulip head of a yukon oct polyaxial screw detached from the screw shank intra-operatively.The broken screw was removed and replaced with another screw of the same size.The surgery was completed successfully after a 30-minute surgical delay.No adverse consequence to the patient was reported.
 
Event or Problem Description
A company representative reported that the tulip head of a yukon oct polyaxial screw detached from the screw shank intra-operatively.The broken screw was removed and replaced with another screw of the same size.The surgery was completed successfully after a 30-minute surgical delay.No adverse consequence to the patient was reported.
 
Additional Manufacturer Narrative
H6 coding has been updated to reflect completion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYAXIAL SCREW; SIZE Ø5.0X30 MM
Common Device Name
POSTERIOR CERVICAL SCREW SYSTEM
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key15759744
Report Number3004774118-2022-00385
Device Sequence Number17698832
Product Code NKG
UDI-Device Identifier10888857350021
UDI-Public10888857350021
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K171444
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7601-05030
Device Catalogue Number7601-05030
Device Lot Number23GA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer01/09/2023
Initial Report FDA Received Date11/09/2022
Supplement Report FDA Received Date02/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR