• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD ANGIOCATH¿ PERIPHERAL VENOUS CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD ANGIOCATH¿ PERIPHERAL VENOUS CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381123
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event is unknown; awareness date has been used for this.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using bd angiocath¿ peripheral venous catheter silicone was visible on the needle.There was no report of patient impact.The following information was provided by the initial reporter: there was potentially any excess silicone present in that lot or if there had been a change in silicone used in the lot.Was lot 1168743 was an aberrant batch or not e.G.Excess silicone".
 
Additional Manufacturer Narrative
H.6.Investigation summary: there was no sample or photo available to bd for evaluation.Therefore, bd was unable to perform a thorough investigation to verify the reported issue.Since, an investigation could not be performed bd was unable to determine a possible root cause.The manufacturing facility has been notified of this incident and the findings.A review of the device history record was performed and no quality issues were found during production.Complaints received for this product and condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported while using bd angiocath¿ peripheral venous catheter silicone was visible on the needle.There was no report of patient impact.The following information was provided by the initial reporter: there was potentially any excess silicone present in that lot or if there had been a change in silicone used in the lot.Was lot 1168743 was an aberrant batch or not e.G.Excess silicone.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ANGIOCATH¿ PERIPHERAL VENOUS CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
Manufacturer (Section G)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15759803
Report Number9610048-2022-00144
Device Sequence Number12810198
Product Code FOZ
UDI-Device Identifier00382903811236
UDI-Public00382903811236
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Catalogue Number381123
Device Lot Number0177306
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date11/09/2022
Supplement Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-