| Model Number |
IPN028484 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device history review could not be conducted since the lot number was not provided.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
Reported issue: the bubbles in the columns are increased which represents for us a pulmonary a/d leak too difficult to evaluate, exa ggerated doubtful bubbles which do not always reflect reality which increases average duration of stay.Additional information indicates that dried blood accumulated at the red/blue connection and there was continuous bubbling.The pleur-evac was changed.The drain was clamped for 24 hours and the chest x-ray shows no subcutaneous emphysema and no pulmonary avulsion, the drain was removed and the patient is ok.The aspiration source was set at 80kpa.
|
| |
|