| Catalog Number |
381311 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Hematoma (1884); Insufficient Information (4580)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the bd insyte-n¿ iv catheter detached from the needle during the perfusion of a premature newborn.The following information was provided by the initial reporter, translated from french: "according to our department, the tip of the catheter became detached from the needle during the perfusion of a premature newborn, causing a haematoma and requiring reperfusion.The offending device was not retained.".
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the bd insyte-n¿ iv catheter detached from the needle during the perfusion of a premature newborn.The following information was provided by the initial reporter, translated from french: "according to our department, the tip of the catheter became detached from the needle during the perfusion of a premature newborn, causing a haematoma and requiring reperfusion.The offending device was not retained.".
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: no photos or samples were received by our quality team for evaluation therefore the failure mode could not be verified.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.Based on the quality team's investigation, the root cause of this incident cannot be determined.
|
| |
|