Model Number ACU0T0
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A other health care professional reported that, during an intraocular lens (iol) implant procedure, the blue injector was stiff and lens would not advance.The procedure was competed by using new iol.There was no patient harm reported.Additional information has been requested.
 
Additional Manufacturer Narrative
The device with the lens was returned in the opened carton.The plunger lock and the lens stop were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was advanced to mid-nozzle and has under-rode the lens.The lens was advanced near the fill line and the optic was misfolded up due to the plunger underride.The trailing haptic was misfolded under the misfolded optic along the left side of the plunger.The leading haptic was extended toward the nozzle tip.The nozzle was removed and cleaned for further evaluation.The lens was removed during cleaning.Top coat dye stain testing was conducted with acceptable results.A qualified viscoelastic was indicated.A plunger underride was observed.The clear lens was located on the plunger in the nozzle entry area.A qualified viscoelastic was indicated.Plunger underride may occur: due to rapid advancement faster than the ifu recommend rate.If inadequate viscoelastic is placed in the device, this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device allowing the plunger to underride the lens.If the device is overfilled with ovd, this can prevent the trailing haptic from being placed properly or move the lens out of position resulting in misfolding.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Top coat dye stain testing was conducted with acceptable results.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15760970
Report Number1119421-2022-02387
Device Sequence Number6239161
Product Code KYB
UDI-Device Identifier00380652395281
UDI-Public00380652395281
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2025
Device Model NumberACU0T0
Device Lot Number15418990
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/17/2022
Supplement Date Received by Manufacturer03/31/2023
Initial Report FDA Received Date11/09/2022
Supplement Report FDA Received Date04/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/03/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient Age82 YR
Patient SexFemale
-
-