| Model Number |
1201-16022 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A company representative reported that a cayman anterior screw, "continued to spin and not lock into the plate," during surgery.The doctor attempted implantation with a second cayman anterior screw but could not get enough purchase in the bone.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.Inspection of the returned devices identified deformation on the threading of the screws.This report captures the first cayman anterior screw.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 coding has been updated to reflect completion of the investigation.
|
| |
|
Event or Problem Description
|
|
A company representative reported that a cayman anterior screw, "continued to spin and not lock into the plate," during surgery.The doctor attempted implantation with a second cayman anterior screw but could not get enough purchase in the bone.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.Inspection of the returned devices identified deformation on the threading of the screws.This report captures the first cayman anterior screw.
|
| |
|